Xuepeng Sun 1,2,3Zhiguo Liu 1,2,3Tianxi Sun 1,2,3,*Longtao Yi 1,2,3[ ... ]Xunliang Ding 1,2,3
Author Affiliations
Abstract
1 The Key Laboratory of Beam Technology and Materials Modification of the Ministry of Education, Beijing Normal University, Beijing 100875, China
2 College of Nuclear Science and Technology, Beijing Normal University, Beijing 100875, China
3 Beijing Radiation Center, Beijing 100875, China
4 Center for Disease Control and Prevention of Beijing, Beijing 100013, China
A tomography device based on a conventional laboratory x ray source, polycapillary parallel x ray lens (PPXRL), and polycapillary collimating x ray lens (PCXRL) is designed. The PPXRL can collect the divergent x ray beam from the source and focus it into a quasi-parallel x ray beam with a divergence of 4.7 mrad. In the center of quasi-parallel x ray beam, there is a plateau region with an average gain in power density of 13.8 and a diameter of 630 μm. The contrast of the image can be improved from 28.9% to 56.0% after adding the PCXRL between the sample and the detector.
340.0340 X-ray optics 220.0220 Optical design and fabrication 
Chinese Optics Letters
2015, 13(9): 093401